Chapter 4
183
ESA Functional Tests
Second Harmonic Spurious Responses
ESA Fu
ncti
onal
T
ests
Second Harmonic Spurious Responses
Test Limits
Test Description
To test second harmonic distortion, a 50 MHz low pass filter is used to filter the source
output, ensuring that harmonics read by the analyzer are internally generated and not
coming from the source. The source power and input attenuation on the analyzer are
adjusted so –20 dBm is the power level at the first mixer.
Model Number
Maximum
E4401B
E7401A
–55 dBc
E4402B
E7402A
–55 dBc
E4403B
–50 dBc
E4404B
E7403A
–55 dBc
E4405B
E7404A
–55 dBc
E4407B
E7405A
–55 dBc
E4408B
–50 dBc
E4411B
–55 dBc
Item
Critical Specifications
(for this test)
Recommended
Agilent Model
Signal Sources
Synthesized Sweeper
8340A/B or 836XX
Series
Miscellaneous
50 MHz Low pass
filter
Rejection at 80 MHz: >60 dB
0955-0306
Adapters
(2) Type-N (m) to
BNC (f)
1250-1476
BNC (f) to BNC (f)
1250-0080