180
Chapter 4
ESA Functional Tests
Scale Fidelity
ESA
Functional T
ests
Scale Fidelity
Test Limits
Test Description
A 50 MHz CW signal is applied from a synthesized sweeper to the input of the analyzer.
The source is adjusted for a response at the reference level. The synthesized sweeper
amplitude is adjusted to achieve a nominal amplitude below the reference level. The
analyzer’s amplitude marker is compared to the actual source change to determine the
scale fidelity error. Most of the error is the source’s output attenuator inaccuracy from the
synthesized sweeper.
dB from Reference Level
Minimum (dB
)
Maximum (dB
)
–4 –1.0
1.0
–16 –1.4
1.4
–28 –1.4
1.4
–40 –1.4
1.4
–52 –1.4
1.4
–64 –2.0
2.0
Item
Critical Specifications
(for this test)
Recommended
Agilent Model
Signal Sources
Synthesized Sweeper
Output Level Accuracy
0 to –15 dBm:
±1
.0 dB
–16 dBm to –63 dBm:
±
1.4 dB
≤
–64 dBm:
≥
2.0 dB
8340A/B or 836XX
Series
Adapters
Type-N (m), to
BNC (f)
1250-1476
Type-N (m), to
APC 3.5 (f)
1250-1475
Cables
Type-N, 152-cm
(60-in)
11500D
BNC, 122-cm (48-in)
10503A
Additional Equipment for 75-Ohm Input
Pad, minimum loss
11852B
Summary of Contents for ESA Series
Page 6: ...6 Contents Table of Contents ...
Page 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Page 33: ...33 Instrument Messages 1 Instrument Messages ...
Page 111: ...111 Functional Tests 2 Functional Tests ...
Page 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Page 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Page 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...