Chapter 3
Making Measurements
Making Return Loss Measurements
5. Make the measurement.
a. Reconnect the device to be measured to the signal separation device.
b. Read the measurement and save it if required.
Example
The following example measures the return loss of a bandpass filter (BPF).
1. Adjust the spectrum analyzer control settings.
With the BPF in the measurement path, adjust the spectrum analyzer control
settings for the correct frequency coverage, resolution bandwidth, input
attenuation and source power.
Having adjusted the control settings, do not alter them during the course of the
measurement.
To obtain a faster sweep, change the coupling from normal spectrum analyzer
mode to stimulus response mode by pressing the
Sweep
front panel key and
Swp Coupling SR SA
menu key until
SR
is underlined.
2. Establish a 0 dB reference trace for normalizing the measured data.
Normalization removes any frequency-response errors introduced by the
components of the test setup before making the measurement. It is performed
by removing the device to be tested and measuring a short or open. As neither
can dissipate the energy of the incident signal—100% reflection takes place,
and the wave is reflected back from the short or open—to the spectrum
analyzer where its value is displayed:
a. Remove the BPF and connect a short in its place as shown in the figure.
b. Press the
View/Trace
front panel key.
c. Press the
More
menu key.
d. Press the
Normalize
menu key,
Store Ref (1 - 3)
and normalize
On.
This procedure establishes a 0 dB reference trace which is stored in the ESA
spectrum analyzer. It is then used to normalize the measured data automatically
by subtracting the short circuit calibration from the measurement obtained with
the device.
Summary of Contents for ESA-E Series
Page 4: ...iv ...
Page 6: ...vi Contents ...
Page 7: ...1 1 Getting Started This chapter describes the process for installing the personality ...
Page 14: ...8 Chapter1 Getting Started Installing the Optional Measurement Personality ...
Page 16: ...10 Chapter2 Menu Maps and Key Descriptions Menus Menus Mode and Measure Menus ...
Page 21: ...15 3 Making Measurements ...
Page 44: ...38 Chapter4 If You Have a Problem Before You Call Agilent Technologies ...
Page 46: ...Index 2 Index ...