Chapter 3
23
Making Measurements
Making Return Loss Measurements
Figure 3-5
Calibrating for Return Loss
* 50 ohm only. The 6 dB attenuators may be used to improve impedance matching.
** 75 ohm only.
NOTE
The distance-to-fault accessory kit, Option B7K, has all the 50 ohm accessories
required to make these measurements.
lists the cables and
accessories required for 50 ohm and 75 ohm measurements.
2. Turn on the tracking generator.
a. Press the
Source
front panel key.
b. Press the
Amplitude
menu key so that the tracking generator is turned on.
c. Set an amplitude level appropriate for the device under test. The default
value = –10 dBm. 0 dBm may be used for systems with higher loss.
3. Adjust the spectrum analyzer control settings (for example frequency,
resolution bandwidth, sweep time and input attenuation) as appropriate for the
signal separation device and device being tested.
4. Establish a 0 dB reference trace for normalizing the measured data.
a. Remove the device to be measured and replace it with a short or open.
b. Press the
View/Trace
front panel key.
c. Press the
More
menu key.
d. Press the
Normalize
menu key,
and normalize
On.
Summary of Contents for ESA-E Series
Page 4: ...iv ...
Page 6: ...vi Contents ...
Page 7: ...1 1 Getting Started This chapter describes the process for installing the personality ...
Page 14: ...8 Chapter1 Getting Started Installing the Optional Measurement Personality ...
Page 16: ...10 Chapter2 Menu Maps and Key Descriptions Menus Menus Mode and Measure Menus ...
Page 21: ...15 3 Making Measurements ...
Page 44: ...38 Chapter4 If You Have a Problem Before You Call Agilent Technologies ...
Page 46: ...Index 2 Index ...