20
Chapter 3
Making Measurements
Making Cable Fault Measurements
c. Press
Calibrate
.
For a 50 ohm setup, connect a load to the analyzer via the power divider as
prompted. See
. For a 75 ohm setup, leave the output port of the
power splitter open.
d. Press
Continue
. Your setup is now calibrated.
Connect a DUT (device under test).
e. Press
Continue
.
Figure 3-3
Cable Fault Location Measurement Setup
* 50 ohm only. The 6 dB attenuators may be used to improve impedance matching.
** 75 ohm only.
NOTE
The distance-to-fault accessory kit, Option B7K, has all the 50 ohm accessories
required to make these measurements.
lists the cables and
accessories required for 50 ohm and 75 ohm measurements.