Installation Note E8364-90022
37
Step 12. Perform Post-Upgrade Adjustments
The following adjustments must be made due to the hardware changes made to the analyzer.
Perform these adjustments in the order listed.
• 3.8 GHz PMYO frequency
• LO power
• offset LO
• source calibration
• phase lock IF gain
• receiver calibration
These adjustments are described in the PNA service guide and in the PNA on-line HELP. A
list of equipment required to perform these adjustments can be found at
Required for Post-Upgrade Adjustments” on page 8
Performance Tests and System Verification
The analyzer should now operate and phase lock over its entire frequency range.
If you experience difficulty with the basic functioning of the analyzer, contact Agilent. Refer to
“Contacting Agilent” on page 3
Although the analyzer functions, its performance relative to its specifications has not been
verified for the additional frequency range enabled by this upgrade.
It is recommended that a full instrument calibration be performed using the PNA
performance test software.
If the testing of the analyzer’s full range of specifications is not required, a system verification
can be performed.
Refer to the analyzer’s service guide for information on performance tests and system
verification.