Chapter 5
261
Language Reference
MMEMory Subsystem
Front Panel
Access:
File, Copy
Move Data to File
:MMEMory:DATA <file_name>,<definite_length_block>
:MMEMory:DATA? <file_name>
Loads
<definite_length_block>
into the memory location
<file_name>
.
The query returns the contents of the <file_name> in the format of a definite length
block. This command can be used for copying files out of the analyzer over the
remote bus. Refer to chapter 3, Programming Examples, for more information.
Example:
Load “abcd” into C:source.txt:
:MMEM:DATA “C:source.txt”,“#14abcd”
Delete a File
:MMEMory:DELete <file_name>
Delete a file.
Example:
:MMEM:DEL “C:source.txt”
Remarks:
If <file_name> does not exist, a “File Name Error” will occur.
Front Panel
Access:
File, Delete
Load a Corrections Table from a File
:MMEMory:LOAD:CORRection
ANTenna|CABLe|OTHer|USER,<file_name>
Loads the data in the file <file_name> to the specified correction set.
Example:
:MMEM:LOAD:CORR ANT, “A:TEST5.CBL”
Front Panel
Access:
File, Load, Type, Corrections
Load a Limit Line from Memory to the Instrument
:MMEMory:LOAD:LIMit LLINE1|LLINE2,<file_name>
Loads a limit line, from the specified file in mass storage to the instrument.
Loading a time limit line deletes any frequency limit lines. Similarly, loading a
frequency limit line deletes any time limit lines.
Example:.
:MMEM:LOAD:LIM LLINE2,“C:mylimit.lim”
Remarks:.
There is no SCPI short form for parameters
LLINE1|LLINE2
.
Front Panel
Access:.
File, Load, Type, Limits