SCPI Command
Displays softkeys for setting measurement parameters (
only for models with Option 413 or 414
). To display
these softkeys, the balanced/unbalanced conversion topology must be set to unbalanced-unbalanced-balanced
SE-SE-Bal
), the balanced/unbalanced conversion function must be on (
BalUn ON
), and the fixture simulator
function must be on (
Fixture Simulator ON
) in the “Analysis Menu” on page 647.
Sss11
Selects parameter S
ss11
. S
ss11
defines the way an unbalanced signal
input to (unbalanced) port 1 on the DUT is reflected as an
unbalanced signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SSS11
Sss21
Selects parameter S
ss21
. S
ss21
defines the way an unbalanced signal
input to (unbalanced) port 1 on the DUT is transmitted to
(unbalanced) port 2 on the DUT as an unbalanced signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SSS21
Sss12
Selects parameter S
ss12
. S
ss12
defines the way an unbalanced signal
input to (unbalanced) port 2 on the DUT is transmitted to
(unbalanced) port 1 on the DUT as an unbalanced signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SSS12
Sss22
Selects parameter S
ss22
. S
ss22
defines the way an unbalanced signal
input to (unbalanced) port 2 on the DUT is reflected as an
unbalanced signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SSS22
Sds31
Selects parameter S
ds31
. S
ds31
defines the way an unbalanced
signal input to (unbalanced) port 1 on the DUT is transmitted to
(balanced) port 3 on the DUT as a differential signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SDS31
Sds32
Selects parameter S
ds32
. S
ds32
defines the way an unbalanced
signal input to (unbalanced) port 2 on the DUT is transmitted to
(balanced) port 3 on the DUT as a differential signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SDS32
Ssd13
Selects parameter S
sd13
. S
sd13
defines the way a differential signal
input to (balanced) port 3 on the DUT is transmitted to
(unbalanced) port 1 on the DUT as an unbalanced signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SSD13
Ssd23
Selects parameter S
sd23
. S
sd23
defines the way a differential signal
input to (balanced) port 3 on the DUT is transmitted to
(unbalanced) port 2 on the DUT as an unbalanced signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SSD23
Scs31
Selects parameter S
cs31
. S
cs31
defines the way an unbalanced
signal input to (unbalanced) port 1 on the DUT is transmitted to
(balanced) port 3 on the DUT as a common mode signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SCS31
Scs32
Selects parameter S
cs32
. S
cs32
defines the way an unbalanced
signal input to (unbalanced) port 2 on the DUT is transmitted to
(balanced) port 3 on the DUT as a common mode signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SCS32
Ssc13
Selects parameter S
sc13
. S
sc13
defines the way a common mode
signal input to (balanced) port 3 on the DUT is transmitted to
(unbalanced) port 1 on the DUT as an unbalanced signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SSC13
Ssc23
Selects parameter S
sc23
. S
sc23
defines the way a common mode
signal input to (balanced) port 3 on the DUT is transmitted to
(unbalanced) port 2 on the DUT as an unbalanced signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SSC23
Sdd33
Selects parameter S
dd33
. S
dd33
defines the way a differential signal
input to (balanced) port 3 on the DUT is reflected as a differential
signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SDD33
Scd33
Selects parameter S
cd33
. S
cd33
defines the way a differential signal
input to (balanced) port 3 on the DUT is reflected as a common
mode signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SCD33
Sdc33
Selects parameter S
dc33
. S
dc33
defines the way a common mode
signal input to (balanced) port 3 on the DUT is reflected as a
differential signal.
:CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB
SDC33
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...