488
Chapter 15
Measurement Examples
Measuring the SAW Bandpass Filter Using the Segment Sweep
2. Create a Segment Sweep Table
Follow the steps below to make entries in the segment sweep table.
Step 1.
Display the segment table.
Setup Description
Key Operation
Presetting
-
OK
Displaying the segment table
-
Edit Segment Table
Step 2.
Display the IF bandwidth setting column in the segment table.
Setup Description
Key Operation
Moving the focus to the softkey menu
Display of IF bandwidth setting column: ON
List IFBW
(turn it
ON
)
NOTE
When setup items (power level, delay time, sweep mode, and sweep time in this case) are
not displayed in the segment table, the setting for the channel in use applies to all segments.
Step 3.
Enter the setup data in the segment table.
Setup Description
Key Operation
Moving the focus to the segment table (select)
Segment 1
Start frequency: 440 MHz
Stop frequency: 915 MHz
Number of points: 47
IF bandwidth: 70 kHz
Segment 2
Start frequency: 915 MHz
Stop frequency: 980 MHz
Number of points: 130
IF bandwidth: 100 kHz
Segment 3
Start frequency: 980 MHz
Stop frequency: 1.035 GHz
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...