Chapter 12
Optimizing Measurements
Reducing Trace Noise
Reducing Trace Noise
Any of the following methods can be used to lower the trace noise.
•
Turning on sweep averaging
•
Turning on smoothing
•
Narrowing the IF bandwidth
For more about sweep averaging and the IF bandwidth, see “Turning on Sweep Averaging”
on page 395 and “Narrowing the IF bandwidth” on page 394.
Turning on Smoothing
Smoothing can be used to reduce noise that has relatively small peaks. By turning on
smoothing, the value of each point on a trace is represented by the moving average over the
values of several nearby points. The smoothing aperture (percentage of sweep span)
defines the range of points to be included in the calculation of the moving average.
NOTE
You can define the smoothing trace by trace.
Figure 12-3
Effects of smoothing (log magnitude format)
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...