Chapter 7
301
Fixture Simulator
Converting Reference Impedance of Balanced Port
7
. F
ixt
ur
e
Sim
ula
to
r
Converting port reference impedance in differential mode
If you turn on the differential port impedance conversion function, the port reference
impedance in the differential mode is converted to an arbitrary value specified with this
function instead of the value in Figure 7-17.
Procedure to Turn On/Off Differential Port Reference Impedance Conversion
Function
Step 1.
- Press
Fixture Simulator
-
Diff Z Conversion
.
Step 2.
Press
Diff Z Conversion
to set the differential impedance conversion function to
ON
.
You can only turn on or off Differential Port Impedance Conversion for all of the balanced
ports together, not for each port individually. If you want to turn off only a specific port, set
the reference impedance of the port to the value in Figure 7-17.
Procedure to Set Differential Port Reference Impedance
Step 1.
- Press
Fixture Simulator
-
Diff Z Conversion
.
Step 2.
Specify the port reference impedance in differential mode.
•
Select the balanced port from
Port 1 (bal) Real
,
Port 2 (bal) Real
, or
Port 3 (bal) Real
to set the differential port reference impedance in “Real” format.
•
Select
Port 1 (bal) Real
and
Port 1 (bal) Imag
,
Port 2 (bal) Real
and
Port2 (bal) Imag
,
or
Port 3 (bal) Real
and
Port 3 (bal) Imag
to set the differential port reference
impedance in “Complex” format.
NOTE
Ports 1, 2, and 3 refer to the logical ports 1, 2, and 3.
Converting port reference impedance in common mode
If you turn on the common port impedance conversion function, the port reference
impedance in the common mode is converted to an arbitrary value specified with this
function instead of the value in Figure 7-17.
Procedure to Turn On/Off Common Port Reference Impedance Conversion
Step 1.
- Press
Fixture Simulator
-
Cmn Z Conversion
.
Step 2.
Press
Cmn Z Conversion
to set the common port impedance conversion function to
ON
.
You can only turn on or off Common Port Impedance Conversion for all of the ports
together, not for each port individually. If you want to turn off only a specific port, set the
impedance of the port to the value in Figure 7-17.
Procedure to set common port reference impedance
Step 1.
- Press
Fixture Simulator
-
Cmn Z Conversion
.
Step 2.
Specify the port reference impedance in common mode.
•
Select the balanced port from
Port 1 (bal) Real
,
Port 2 (bal) Real
, or
Port 3 (bal) Real
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...