Chapter 6
257
Data Analysis
Analyzing Data on the Trace Using the Marker
6. Da
ta
Analysis
Reading only the actual measurement point/Reading the value
interpolated between measurement points
The point on the trace on which a marker can be placed differs depending on how the
discrete marker mode is set up.
Turning on discrete
mode
(
Discrete ON
)
A marker moves only between actual measurement points. When a
specific marker stimulus value is specified as a numerical value, the
marker is placed at the measurement point closest to the specified
value. A marker that is placed between interpolated points with the
discrete mode off automatically moves to the nearest measurement
point when the discrete mode is turned on.
Turning off discrete
mode
(
Discrete OFF
)
The marker can move from one actual measurement point to
another. Because it is interpolated, it can also move in the space
between measurement points.
Figure 6-2
Marker discrete mode
Turning Discrete Mode On or Off
Step 1.
Press
(or
) and
(or
) to activate the trace on
which the discrete mode is set up.
Step 2.
Press
.
Step 3.
Press
Discrete
to turn the discrete mode on or off.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...