![Agilent Technologies E5070B User Manual Download Page 240](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065240.webp)
234
Chapter 4
Calibration
Scalar-Mixer Calibration
Figure 4-74
frequency-offset error model
Confirming calibration status
Error correction status of each channel
You can confirm the progress of error correction for each channel by viewing the error
calibration status.
The error correction status is displayed in the channel status bar in the lower part of the
window by one of the following symbols:
Symbol
Execute status of error correction
Cor
(displayed in
blue)
Error Correction: On (enabled for all traces)
Cor
(displayed in
gray)
Error Correction: On (enabled for some traces)
Off
(displayed in gray)
Error Correction: Off
---
(displayed in gray)
Error correction: On (no calibration data)
C?
(displayed in blue)
Error correction: On (interpolation is being executed or the IF
bandwidth, sweep delay time, power level, or sweep type is different
from when calibration was executed.)
C!
(displayed in blue)
Error correction: On (extrapolation is being executed.)
For the traces in which the scalar-mixer is valid, the status [S2] is added (Figure 4-75).
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...