![Agilent Technologies E5070B User Manual Download Page 187](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065187.webp)
Chapter 4
181
Calibration
Changing the Calibration Kit Definition
4. Cal
ibra
tion
Changing the Calibration Kit Definition
In most measurements, the user can use pre-defined calibration kits as they are. However, it
may be necessary to change the definition of a calibration kit (or create a new one) when
changing the pre-defined connector between male and female (e.g. from OPEN (f) to
OPEN (m)
*1
) or when a special standard is used or a high degree of accuracy is demanded.
When it is necessary to change the definition of a calibration kit that contains a calibration
device but no calibration kit model, the user must fully understand error correction and the
system error model.
A user-defined calibration kit may be used in the following circumstances.
•
When the user wants to use connectors other than those pre-defined in the calibration
kits for the E5070B/E5071B (e.g., a SMA connector).
•
When the user wants to use different standards in place of one or more standards
pre-defined in the E5070B/E5071B. For example, when three offset SHORT standards
are used instead of OPEN, SHORT, and LOAD standards.
•
When the user wants to modify the standard model of a pre-defined calibration kit and
turn it into a more accurate model. It is possible to perform better calibration if the
performance of the actual standard is better reflected in the standard model. For
example, you may need to define the 7-mm LOAD standard as 50.4
Ω
instead of 50.0
Ω
.
*1.The designations (m) and (f) in the name (label) of the standard indicate male (m) and female
(f) analyzer connectors.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...