Chapter 4
161
Calibration
Adapter Characterization
4. Cal
ibra
tion
Concept
Adapter characterization is a function that calculates the S-parameters of an adapter based
on 3 measurement results obtained by using open/short/load standards connected to the test
port, via the adapter, for which 1-port calibration has been performed. This VBA macro
uses test port 1 for this measurement.
The S-parameters of the adapter can be calculated from the above 3 measurement results.
However, you can calculate the S-parameters more accurately by connecting the adapter to
the test port in the reverse direction, measuring 3 more results in the same way as above,
and using the above 3 results with the 3 new results (a total of 6 measurement results). This
VBA macro uses test port 2 for the measurement made when the adapter is connected in
the reverse direction.
Figure 4-40
Adapter characterization
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...