Chapter 2
19
Installation
Environmental Requirements
2. Ins
tall
ati
on
Ventilation Requirements
To ensure safety requirements, the specifications and measurement accuracy of the
analyzer, you must keep environmental temperature within the specified range by
providing appropriate cooling clearance around the analyzer or, for the rackmount type, by
forcefully air-cooling inside the rack housing. For more information on environmental
temperature to satisfy the specifications and measurement accuracy of the analyzer, see
“Specifications And Supplemental Information“ in the
E5070A/E5071A User’s Guide
.
When the environmental temperature around the analyzer is kept within the temperature
range of the operating environment specification (See Section “Operating Environments”
on page 18 ), the analyzer conforms to the requirements of the safety standard.
Furthermore, under that temperature requirement, the analyzer still conforms to the
requirements of the safety standard even when placing the analyzer with cooling clearance
as follows:
Figure 2-2
Ventilation space at the installation
Requirements
Back
³
180 mm
Sides
³
0 mm (both right and left)
Summary of Contents for E5070A ENA Series
Page 12: ...12 Chapter1 Introduction of This Manual Overview of the E5070A E5071A ...
Page 30: ...30 Chapter2 Installation Power Supply and Blown Fuses Figure 2 10 Power cable options ...
Page 56: ...56 Chapter3 Quick Start Guide Measurement Example of a Bandpass Filter ...
Page 61: ...Index 61 Index U USB port front panel 41 User s Guide 6 V VBA Programmer s Guide 6 ...
Page 62: ......