Agilent Technologies E4430B Installation Note Download Page 1

Part Number E4400-90167

Printed in USA

April 2000 Supersedes August 1999

Installation Note

Agilent Technologies ESG-D Series Signal Generators

Digital Upgrade Kit (Option UN3 to UN8)

Kit Part Number E4400-60160

Models: E4430B, E4431B, E4432B and E4433B

Summary of Contents for E4430B

Page 1: ... Printed in USA April 2000 Supersedes August 1999 Installation Note Agilent Technologies ESG D Series Signal Generators Digital Upgrade Kit Option UN3 to UN8 Kit Part Number E4400 60160 Models E4430B E4431B E4432B and E4433B ...

Page 2: ...d to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Copyright 1999 2000 Agilent Technologies ...

Page 3: ...ignal generator 2 removing Option UN3 3 installing Option UN8 4 installing and verifying new firmware 5 activating Option UN8 6 recalibrating and reassembling the signal generator Product Affected E4430B E4431B E4432B and E4433B ESG D Series Signal Generators Serial Numbers All Options UN3 Compatibilities To Be Performed By X Agilent Technologies Service Center X Personnel Qualified by Agilent Tec...

Page 4: ...rkstation Item Quantity Description Part Number Std No Opt s Implies no other digital hardware options are installed U8 to UN7 U8 to UND 1 1 1 1 Board Assembly Real Time I Q Baseband Generator A7 E4400 60070 2 1 1 1 Board Assembly Generator 1 Meg A8 E4400 60154 3 1 0 0 Firmware Upgrade Kit Digital E4400 60230 4 1 1 1 UN8 Option Label 7120 1232 5 1 1 1 Ribbon Cable Interconnect W19 8120 8349 6 1 1 ...

Page 5: ...eue by pressing Utility Error Info The first error message in the queue is shown in the text area of the display Refer to the service guide for information about the error message If there is more than one error message each message will be designated as 1 of n press the View Next Error Message softkey until you have seen all of the messages 5 When you have resolved all of the error messages press...

Page 6: ...N3 Hardware 1 Turn off the line power switch on the signal generator and disconnect the line cord 2 Loosen the two screws on each handle and remove the two strap handles Figure 1 page 7 items 1 and 2 3 Remove the four bottom feet Figure 1 item 3 4 Remove the four rear feet Figure 1 items 4 and 5 5 Slide the top cover off the signal generator Figure 1 item 6 6 Remove the 11 top cover screws and rem...

Page 7: ...Installation Note E4400 90167 7 Figure 1 Cover Removal Figure 2 Option UN3 A7 Baseband Generator Board and Cables ...

Page 8: ...8 Installation Note E4400 90167 Figure 3 Option UND A5 Dual Arb Board and Cables ...

Page 9: ...Installation Note E4400 90167 9 Figure 4 Option UN7 A6 Bit Error Rate Test Board and Cables Figure 5 Option UN3 A8 Data Generator Board and Cables ...

Page 10: ...to both A8P2 and A5P1 7 Connect W17 I Out to the rear panel to A7P405 NOTE If Option UND is installed reconnect W17 to A5J1 instead of A7P405 8 Connect W16 Q Out to the rear panel to A7P404 NOTE If Option UND is installed reconnect W16 to A5J2 instead of A7P404 9 Connect W15 BASEBAND GENREF IN to the rear panel to A7P403 NOTE If Option UND is installed reconnect W15 to A5J3 instead of A7P403 Then ...

Page 11: ...Installation Note E4400 90167 11 Figure 6 Option UN8 Top View Figure 7 Cables W25 and W26 ...

Page 12: ...el to A7P403 Q OUT W16 9 Rear Panel to A7P404 I OUT W17 09 Rear Panel to A7P405 DATA W3 5 Front Panel to A14P5 SYMBOL SYNC W5 6 Front Panel to A14P7 DATA CLOCK W4 7 Front Panel to A14P6 Rear Panel Interface Cable W18 Rear Panel Interface A17 to A8P2 INT Q W25 05 Daughter Board A15 to A14P103 INT I W26 06 Daughter Board A15 to A14P102 Option UN8 Option UN7 Option UN7 Connections Figure 4 on page 9 ...

Page 13: ...Installation Note E4400 90167 13 Figure 8 Options UN8 UN7 UND Top View ...

Page 14: ...nable the selected options press Proceed With Reconfiguration then press Confirm Change Instrument will Reboot CAUTION If you enable an option without the required hardware installed only the menus for that option are activated the option does not function Internally Recalibrating the Signal Generator 1 Preset the signal generator then press the MENUS hardkey Utility 2 In the Utility softkey menu ...

Page 15: ...k on the signal generator Figure 1 item 6 4 Replace the four rear feet and torque the four screws to 21 in lb Figure 1 items 4 and 5 5 Replace the two strap handles one on each side of the instrument and torque the two screws on each handle to 21 in lb Figure 1 and 2 6 Replace the four bottom feet Figure 1 item 3 7 Cut out a single UN8 label and attach it over the UN3 serial tag option located on ...

Page 16: ...16 Installation Note E4400 90167 ...