Chapter 4
179
Basic Digital Operation
Modifying a FIR Filter Using the FIR Table Editor
Modifying the Coefficients
1. Using the front-panel arrow keys, highlight coefficient 15.
2. Press
0
>
Enter
.
3. Press
Display Impulse Response
.
Figure 4-35
Refer to
. The graphic display can provide a useful troubleshooting tool (in this
case, it indicates that a coefficient value is missing, resulting in an improper Gaussian response).
4. Press
Return
.
5. Press
More (2 of 2)
.
6. Highlight coefficient 15.
7. Press
1
>
Enter
.
Storing the Filter to Memory
The maximum file name length is 23 characters (alphanumeric and special characters).
1. Press
Load/Store
>
Store To File
.
2. Name the file NEWFIR2.
3. Press
Enter
.
The contents of the current FIR table editor are stored to a file in non-volatile memory and the catalog of FIR
files is updated to show the new file.
Summary of Contents for E4428C
Page 22: ...Contents xxii ...
Page 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Page 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Page 229: ...205 6 Analog Modulation ...
Page 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Page 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Page 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Page 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Page 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Page 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Page 667: ...643 18 Troubleshooting ...
Page 700: ...Index 676 Index ...