Chapter 3
81
Basic Operation
Using Security Functions
User IQ Cal File (Vector Models Only)
When a user-defined IQ calibration has been performed, the cal file data is removed by setting the cal file to default, as
follows:
1.
On the front panel, press:
I/Q
>
I/Q Calibration
>
Revert to Default Cal Settings
2.
Send these commands:
•
:CAL:IQ:DEF
•
:CAL:WBIQ:DEF
Using the Secure Mode
The secure mode automatically applies the selected
Security Level
action the next time the instrument’s
power is cycled.
Setting the Secure Mode Level
CAUTION
Do not use the functions
Erase
,
Overwrite
, or
Sanitize
with signal generator firmware
revisions C.04.84, C.04.86, or C.04.95. If you have one of these revisions installed, please
upgrade
immediately
to revision C.04.96 or later.
1. Press
Utility
>
Memory Catalog
>
More (1 of 2)
>
Security
>
Security Level
.
2. Choose from the following selections:
None
−
equivalent to a factory preset, no user information is lost
Erase
−
equivalent to
Erase All
Overwrite
−
equivalent to
Erase and Overwrite All
Sanitize
−
equivalent to
Erase and Sanitize All
Activating the Secure Mode
CAUTION
Once you activate secure mode (by pressing
Confirm)
, you cannot deactivate or decrease the
security level; the erasure actions for that security level execute at the next power cycle.
Once you activate secure mode, you can only increase the security level until you cycle
power. For example, you can change
Erase
to
Overwrite
, but not the reverse.
After the power cycle, the security level selection remains the same, but the secure mode is
not activated.
Press
Utility
>
Memory Catalog
>
More (1 of 2)
>
Security
>
Enter Secure Mode
>
Confirm
.
Summary of Contents for E4428C
Page 22: ...Contents xxii ...
Page 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Page 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Page 229: ...205 6 Analog Modulation ...
Page 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Page 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Page 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Page 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Page 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Page 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Page 667: ...643 18 Troubleshooting ...
Page 700: ...Index 676 Index ...