Power Meter Operation
2
E4419B Power Meter User’s Guide
103
confidence check has been successful. While the test is running, the
message “
Testing...
” appears. If the correct power meter reading is
made, the message “
Passed
” is displayed, otherwise “
Failed
” is displayed.
If the confidence check fails, the failure appears in the error queue. To
examine the error queue, go to the “Errors” screen (refer to
).
Remote Testing
To invoke the remote self test, the IEEE 488.1 compliant standard command,
*TST?
is used. This command runs a full self test and returns one of the
following codes:
•
0 - no tests failed
•
1 - one or more tests failed
The remote self test consists of the following tests:
•
ROM checksum
•
RAM
•
RAM Battery
•
Display Assembly
•
Calibrator
•
Measurement Assemblies
•
Communications Assembly (Implicit)
The communications assembly is tested implicitly, in that the command will
not be accepted or return a result unless the GPIB interface is functioning
correctly.
Refer to
on page 104, if you require a description of each
individual test.
When the
*TST?
command is executed, the screen is cleared. As each test
takes place, the name of the test is listed on the screen. While a test is running,
the message “
Testing...
” appears next to the name of the test. As each
stage of the test is completed, the message “
Testing...
” is replaced by
either the message “
Passed
” or “
Failed
”.
Summary of Contents for E4419B
Page 1: ...Agilent Technologies Agilent E4419B Power Meter User s Guide ...
Page 14: ...xiv E4419B Power Meter User s Guide ...
Page 22: ...xxii E4419B Power Meter User s Guide ...
Page 117: ...Power Meter Operation 2 E4419B Power Meter User s Guide 95 Figure 2 22 Save Recall Screen ...
Page 136: ...2 Power Meter Operation 114 E4419B Power Meter User s Guide ...
Page 174: ...3 Menu Reference 152 E4419B Power Meter User s Guide ...
Page 184: ...4 Error Messages 162 E4419B Power Meter User s Guide ...