![Agilent Technologies E4418B User Manual Download Page 145](http://html.mh-extra.com/html/agilent-technologies/e4418b/e4418b_user-manual_2868035145.webp)
Menu Reference
3
E4418B Power Meter User’s Guide
121
The softkey menu structured to this key affects the channel setup.
Press this key to access the “Frequency/Cal Fac” menu which allows you to
enter the frequency of the signal to be measured or the calibration factor.
Depending on the type of power sensor connected and whether or not a sensor
calibration table or frequency dependent offset table is selected, the text on
some softkeys may appear grayed out. This indicates that they are not relevant
in the power meter’s current mode of operation.
details the
relevant softkeys in various modes of operation.
Table 3-1
Combination of Sensor Type and Correction Applied to Current Measurement
Sensor Model
Sensor Calibration Table Selected
Sensor Calibration Table Not Selected
Frequency
Cal Factor
Frequency
Cal Factor
Agilent 8480 Series
power sensors
Frequency can be
entered. Default value
is 50 MHz.
The Cal Factor is
obtained from the
sensor calibration
table but can be
overridden by this
softkey.
Frequency can be
entered if a
frequency
dependent offset
table is selected.
The Cal Factor can be
entered. Default
value is 100%.
Agilent E-Series
power sensors
Not Applicable
Not Applicable
Frequency can be
entered. Default
value is 50 MHz.
The Cal Factor
cannot be entered, it
is obtained from the
EEPROM in the
power sensor.
Agilent N8480 Series
power sensors
(excluding Option
CFT)
Not Applicable
Not Applicable
Frequency can be
entered. Default
value is 50 MHz.
The Cal Factor
cannot be entered, it
is obtained from the
EEPROM in the
power sensor.
Agilent N8480 Series
power sensors with
Option CFT
Frequency can be
entered. Default value
is 50 MHz.
The Cal Factor is
obtained from the
sensor calibration
table but can be
overridden by this
softkey.
Frequency can be
entered if a
frequency
dependent offset
table is selected.
The Cal Factor can be
entered. Default
value is 100%.
Freq
Cal Fac
Freq
Cal Fac
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...