Specifications
Power Meter Supplemental Characteristics
1-8
Agilent E4418B/E4419B Service Guide
Table 1-3: Power Sensor Specifications±
Power Sensor
Zero Drift
1
1. Within 1 hour after zero set, at a constant temperature, after a 24 hour
warm-up of the power meter.
Measurement Noise
2
2. The number of averages at 16 (for normal mode) and 32 (for x2 mode), at
a constant temperature, measured over a 1 minute interval and 2 standard
deviations. For Agilent E-series power sensors the measurement noise is
measured within the low range. Refer to the relevant power sensor manual
for further information.
Agilent 8481A
<
±
10 nW
<
110 nW
Agilent 8481B
<
±
10
µ
W
<
110
µ
W
Agilent 8481D
<
±
4 pW
<
45 pW
Agilent 8481H
<
±
1
µ
W
<
10
µ
W
Agilent 8482A
<
±
10 nW
<
110 nW
Agilent 8482B
<
±
10
µ
W
<
110
µ
W
Agilent 8482H
<
±
1
µ
W
<
10
µ
W
Agilent 8483A
<
±
10 nW
<
110 nW
Agilent 8485A
<
±
10 nW
<
110 nW
Agilent 8485D
<
±
4 pW
<
45 pW
Agilent R8486A
<
±
10 nW
<
110 nW
Agilent R8486D
<
±
6 pW
<
65 pW
Agilent Q8486A
<
±
10 nW
<
110 nW
Agilent Q8486D
<
±
6 pW
<
65 pW
Agilent W8486A
<
±
40 nW
<
450 nW
Agilent 8487A
<
±
10 nW
<
110 nW
Agilent 8487D
<
±
4 pW
<
45 pW
Agilent E4412A
<±
15 pW
<
70 pW
Agilent E4413A
<±
15 pW
<
70 pW
4402serv.book Page 8 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...