100
Chapter 4
Making Measurements
Making the Waveform (Time Domain) Measurement
Changing the Measurement Setup
NOTE
Parameters that are under the
Advanced
key seldom need to be
changed. Any changes from the default values may result in invalid
measurement data.
Table 2
Waveform (Time Domain) Measurement Defaults
Measurement Parameter
Factory Default Condition
View/Trace
RF Envelope
Sweep Time
2.000 ms
Res BW
100.000 kHz
Averaging:
Avg Number
Avg Mode
Avg Type
10 Off
Exp
Pwr Avg (RMS)
Trigger Source
Free Run (Immediate)
RF Envelope Window:
Amplitude Y Scale
Scale/Div
Reference
10.00 dB
0.00 dBm (Top)
I/Q Waveform Window:
Amplitude Y Scale
Scale/Div
Reference
100.0 mv
0.00 V (Ctr)
Advanced
Pre-ADC BPF
Off
RBW Filter
Gaussian
ADC Range
Auto
Data Packing
Auto
ADC Dither
Off
Decimation
Off