
Chapter 5
371
Language Reference
TRIGger Subsystem
Automatic Trigger Time
:TRIGger[:SEQuence]:AUTO[:TIME] <time>
:TRIGger[:SEQuence]:AUTO[:TIME]?
After the measurement is activated the instrument will take a data
acquisition immediately upon receiving a signal from the selected
trigger source. If no trigger signal is received by the end of the time
specified in this command, a data acquisition is taken anyway.
TRIGger[:SEQuence]:AUTO:STATE must be on.
Factory Preset
and *RST:
100.0 ms
Range:
1.0 ms to 1000.0 s
0.0 to 1000.0 s for cdma2000, W-CDMA (3GPP),
W-CDMA (Trial & ARIB)
Default Unit:
seconds
Front Panel External Trigger Delay Value
:TRIGger[:SEQuence]:EXTernal[1]:DELay <time>
:TRIGger[:SEQuence]:EXTernal[1]:DELay?
Set the amount of trigger delay when using the front panel external
trigger input. Set the trigger value to zero (0) seconds to turn off trigger
delay.
Factory Preset
and *RST:
0.0 s
Range:
−
500.0 ms to 500.0 ms
−
100.0 ms to 500.0 ms for cdma2000, W-CDMA (3GPP),
W-CDMA (Trial & ARIB)
Default Unit:
seconds
Front Panel
Access:
Mode Setup, Trigger, Ext Front, Delay
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...