Chapter 5
335
Language Reference
SENSe Subsystem
Spectrum—Frequency Span
[:SENSe]:SPECtrum:FREQuency:SPAN <freq>
[:SENSe]:SPECtrum:FREQuency:SPAN?
Set the frequency span to be measured.
Factory Preset
and *RST:
1.0 MHz
100.0 kHz for iDEN mode
Range:
10 Hz to 10.0 MHz (15 MHz when Service mode is
selected)
Default Unit:
Hz
Remarks:
The actual measured span will generally be slightly
wider due to the finite resolution of the FFT.
To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Spectrum—Sweep (Acquisition) Time
[:SENSe]:SPECtrum:SWEep:TIME <time>
[:SENSe]:SPECtrum:SWEep:TIME?
Set the sweep (measurement acquisition) time. It is used to specify the
length of the time capture record. If the specified value is less than the
capture time required for the specified span and resolution bandwidth,
the value is ignored. The value is set at its auto value when auto is
selected. This is an advanced control that normally does not need to be
changed.
Factory Preset
and *RST:
188.0
µ
s
15.059 ms, for iDEN mode
Range:
100 ns to 10 s
Default Unit:
seconds
Remarks:
NOTE: You must be in the Service mode to use this
command. Use INSTrument:SELect to set the mode.
This command only effects the RF envelope trace.
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...