214
Chapter 5
Language Reference
DISPlay Subsystem
Spectrum - Y-Axis Reference Level
:DISPlay:SPECtrum[n]:WINDow[m]:TRACe:Y[:SCALe]:RLEVel
<power>
:DISPlay:SPECtrum[n]:WINDow[m]:TRACe:Y[:SCALe]:RLEVel?
Sets the amplitude reference level for the y-axis.
n
−
selects the view, the default is Spectrum.
— n=1, Spectrum
— n=2, I/Q Waveform
— n=3, numeric data (service mode)
— n=4, RF Envelope (service mode)
m
−
selects the window within the view. The default is 1.
Factory Preset
and *RST:
0 dBm, for Spectrum
Range:
−
250 to 250 dBm, for Spectrum
Default Unit:
dBm, for Spectrum
Remarks:
May affect input attenuator setting.
To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Front Panel
Access:
When in Spectrum measurement:
Amplitude Y Scale, Ref
Level