204
Chapter 5
Language Reference
CALibration Subsystem
Load the Factory Default Calibration Constants
:CALibration:LOAD:DEFault
Load the factory default alignment data, ignoring the effect of any
alignments already done.
Front Panel
Access:
System, Alignments, Restore Align Defaults
Align the Wide LC Prefilter
:CALibration:PFILter:LCWide
:CALibration:PFILter:LCWide?
Align the wide LC prefilter. (1.2 MHz to 7.5 MHz)
Remarks:
A valid service password needs to be entered prior to
sending the command.
Front Panel
Access:
System, Diagnostics
Align the Narrow LC Prefilter
:CALibration:PFILter:LCNarrow
:CALibration:PFILter:LCNarrow?
Align the narrow LC prefilter. (200 kHz to 1.2 MHz)
Remarks:
A valid service password needs to be entered prior to
sending the command.
Front Panel
Access:
System, Alignments, Align Subsystem, IF
Align the Wide Crystal Prefilter
:CALibration:PFILter:XTALWide
:CALibration:PFILter:XTALWide?
Align the wide crystal prefilter. (20 kHz to 200 kHz)
Remarks:
A valid service password needs to be entered prior to
sending the command.
Front Panel
Access:
Enter service password and press
System, Diagnostics
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...