![Agilent Technologies E364XA Series User Manual Download Page 97](http://html1.mh-extra.com/html/agilent-technologies/e364xa-series/e364xa-series_user-manual_2868438097.webp)
Chapter 4 Remote Interface Reference
Calibration Commands
96
Calibration
Example
1 Select the output to be calibrated and enable the output of the power
supply by sending the commands:
‘‘
INST:SEL {OUT1|OUT2}
’’
‘‘
OUTP ON
’’
2 Disable the voltage protection function.
‘‘
VOLT:PROT:STAT OFF
’’
3 Unsecure the power supply with the secure code before calibration.
‘‘
CAL:SEC:STAT OFF,
‘
<code>
’’’
4 For voltage calibration, connect a digital voltmeter (DVM) across the
power supply’s output terminals.
5 Set the power supply to low-end (MIN) calibration point.
‘‘
CAL:VOLT:LEV MIN
’’
6 Enter the reading you obtained from the DVM.
‘‘
CAL:VOLT:DATA 0.549
’’
7 Set the power supply to middle (MID) calibration point.
‘‘
CAL:VOLT:LEV MID
’’
8 Enter the reading you obtained from the DVM.
‘‘
CAL:VOLT:DATA 11.058
’’
9 Set the power supply to high (MAX) calibration point.
‘‘
CAL:VOLT:LEV MAX
’’
10 Enter the reading you obtained from the DVM.
‘‘
CAL:VOLT:DATA 21.566
’’
11 Set the power supply to overvoltage protection calibration point.
‘‘
CAL:VOLT:PROT
’’
12 For current calibration, connect an appropriate current monitoring
resistor (shunt) across the output terminals and connect the DVM
across the shunt resistor.
13 Repeat steps (5) through (9) by substituting ‘‘
CURR
’’ for ‘‘
VOLT
’’ for
current calibration. For example, ‘‘
CAL
:CURR
:LEV MIN
’’
14 Repeat steps (1) through (13) for the other output calibration.
15 Record calibration information such as next calibration due date or
contact person for future reference. The calibration string may contain
up to 40 characters.
‘‘
CALibration:STRing
‘
<string>
’’’
N o t i c e :
You should wait for the DVM reading to stabilize for accurate calibration.
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......