Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
86
VOLTage:PROTection {
<voltage>
|MINimum|MAXimum}
Set the voltage level at which the overvoltage protection (OVP) circuit will trip.
If the peak output voltage exceeds the OVP level, then the power supply output
is shorted by an internal SCR. An overvoltage condition can be cleared with
the
VOLT:PROT:CLE
command after the condition that caused the OVP trip
is removed.
VOLTage:PROTection? [MINimum | MAXimum]
Query the overvoltage protection trip level presently programmed.
VOLTage:PROTection:STATe {0 | 1 | OFF | ON}
Enable or disable the overvoltage protection function. At
*RST
, this value is
set to ‘‘ON’’.
VOLTage:PROTection:STATe?
Query the state of the overvoltage protection function. The returned parameter
is ‘‘0’’ (OFF) or ‘‘1’’ (ON).
VOLTage:PROTection:TRIPped?
Return a ‘‘1’’ if the overvoltage protection circuit is tripped and not cleared or
a ‘‘0’’ if not tripped.
VOLTage:PROTection:CLEar
Cause the overvoltage protection circuit to be cleared. After this command,
the output voltage is restored to the state it was in before the protection feature
occurred and the OVP trip level remains unchanged to the value presently
programmed. Before sending this command, lower the output voltage below
the trip OVP point, or raise the OVP trip level above the output setting.
Note
that the overvoltage condition caused by an external source must be removed
first before proceeding this command.
VOLTage:RANGe {P8V* | P20V* | P35V** | P60V** | LOW | HIGH}
Select an output range to be programmed by the identifier. For example,
‘‘P20V’’ or ‘‘HIGH’’ is the identifier for the 20V/1.5A range and ‘‘P8V’’ or ‘‘LOW’’
is for the 8V/3A range (for E3646A model). At
*RST
, low voltage range is
selected.
VOLTage:RANGe?
Query the currently selected range. The returned parameter is ‘‘P8V’’ or ‘‘P35V’’
for low voltage range, or ‘‘P20V’’ or ‘‘P60V’’ for high voltage range.
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......