Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
82
CURRent:STEP? [DEFault]
Return the value of the step size currently specified. The returned parameter
is a numeric value. ‘‘
DEFault
’’ gives the minimum resolution of the step size
in unit of amps.
CURRent:TRIGgered {
<current>
| MINimum | MAXimum}
Program the pending triggered current level. The pending triggered current
level is a stored value that is transferred to the output terminals when a trigger
occurs. A pending triggered level is not affected by subsequent
CURRent
commands.
CURRent:TRIGgered? [MINimum | MAXimum]
Query the triggered current level presently programmed. If no triggered level
is programmed, the
CURRent
level is returned.
Example
The following program segments show how to use the
CURR
UP
or
CURR
DOWN
command to increase or decrease the output current with the
CURR
:
STEP
command.
‘‘
CURR
:
STEP
0.01
’’
Set the step size to 0.01 A
‘‘
CURR
UP
’’
Increase the output current
‘‘
CURR
:
STEP
0.02
’’
Set the step size to 0.02 A
‘‘
CURR
DOWN
’’
Decrease the output current
INSTrument[:SELect] {OUTPut1 | OUTPut2 | OUT1 | OUT2}
Select the output to be programmed one of the two outputs by the output
identifier. The outputs of the power supply are considered as two logical
instruments. The
INSTrument
command provides a mechanism to identify
and select an output. When one output is selected, the other output is
unavailable for programming until selected. The commands which are affected
by the
INSTrument
command are output setting command (
SOURce
),
measurement command (
MEASure
), and calibration command
(
CALibration
).
INSTrument[:SELect]?
Return the currently selected output by the
INSTrument{:SELect]
or
INSTrument:NSELect
command. The returned value is ‘‘OUTP1’’ or
‘‘OUTP2’’.
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......