Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
81
4
Output Setting and Operation Commands
This section describes low-level commands used to program the power supply.
Although the
APPLy
command provides the most straightforward method to
program the power supply, the low-level output setting commands give you
more flexibility to change the individual parameters.
CURRent{
<current>
| MINimum | MAXimum | UP | DOWN}
Program the immediate current level of the power supply. The immediate level
is the current value of the output terminals.
The
CURRent
command changes the output of the power supply to the newly
programmed value regardless of the output range presently selected.
You can substitute ‘‘
MINimum
’’ or ‘‘
MAXimum
’’ in place of a specific value for
the current parameter.
MIN
selects the lowest current values of ‘‘0’’ amps.
MAX
selects the highest current values allowed for the selected range.
This command also increases or decreases the immediate current level using
the ‘‘
UP
’’ or ‘‘
DOWN
’’ parameter by a predetermined amount. The command
CURRent
:
STEP
sets the amount of increase or decrease.
Notice
that a new
increment setting will cause an execution error -222 (Data out of range) when
the maximum or the minimum rated current is exceeded.
CURRent? [MINimum | MAXimum]
Return the presently programmed current level of the power supply.
CURR?
MAX
and
CURR? MIN
return the highest and lowest programmable
current levels for the selected range.
CURRent:STEP {
<numeric value>
| DEFault}
Set the step size for current programming with the
CURRent
UP
and
CURRent
DOWN
commands. See the example in the
next
page.
To set the step size to the minimum resolution, set the step size to ‘‘
DEFault
’’.
The minimum resolution of the step size is approximately 0.052 mA (E3646A),
0.014 mA (E3647A), 0.095 mA (E3648A), and 0.027 mA (E3649A), respectively.
The
CURR:STEP? DEF
returns the minimum resolution of your instrument.
The immediate current level increases or decreases by the value of the step
size. For example, the output current will increase or decrease 10 mA if the
step size is 0.01. At
*RST
, the step size is the value of the minimum resolution.
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......