![Agilent Technologies E364XA Series User Manual Download Page 66](http://html1.mh-extra.com/html/agilent-technologies/e364xa-series/e364xa-series_user-manual_2868438066.webp)
Chapter 3 Front-Panel Operation and Features
Calibration Overview
65
3
To Unsecure for Calibration
You can unsecure the power supply either from the front panel or over the
remote interface. The power supply is secured when shipped from the factory.
See the table 3-1 for the factory setting secure code for your power supply.
1 Select the calibration mode.
If the power supply is secured, the above message is displayed as you turn on
the power supply by holding down
(
Calibrate
) key and hold down the
key
until you hear a long beep
. And a message ‘‘
CAL MODE
’’ is displayed.
2 Enter the security code.
Enter the security code using the control knob and resolution selection keys.
3 Save the change and exit the menu.
You will see the above message if the security code is correct. And a message
‘‘
CAL MODE
’’ is displayed. To exit the calibration mode, turn the power off
and on.
Notice that if you enter the wrong secure code, “
INVALID
” is displayed and
the code entering mode is displayed for you to enter the correct code.
•
Remote Interface Operation:
CAL:SEC:STAT {OFF|ON},<code>
Secure or unsecure the power supply
To unsecure the power supply, send the above command with the same
code used to secure. For example,
‘‘
CAL:SEC:STAT OFF,
‘
003646
’’’
(E3646A model)
SECURED
000000
UNSECURED
Power
Calibrate
View
Calibrate
View
Secure
Config
I/O
Secure
Config
I/O
Power
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......