![Agilent Technologies E364XA Series User Manual Download Page 44](http://html1.mh-extra.com/html/agilent-technologies/e364xa-series/e364xa-series_user-manual_2868438044.webp)
Chapter 3 Front-Panel Operation and Features
Constant Current Operation
43
3
5 Adjust the knob for the desired output current.
Check that the
Limit
annunciator still flashes. Set the knob for
current
control.
Change the flashing digit using the resolution selection keys and adjust the
knob to the desired output current.
6 Return to the meter mode.
Press
or let the display time-out after several seconds to return to the
meter mode. Notice that the
Limit
annunciator turns off and the display shows
“
OUTPUT OFF
” message.
7 Enable the output.
The
OFF
annunciator turns off and the
CC
annunciator turns on. Notice that
the display is in the
meter
mode.
8 Verify that the power supply is in the constant current mode.
If you operate the power supply in the constant current (CC) mode, verify that
the
CC
annunciator is lit. If the
CV
annunciator is lit, choose a higher voltage
limit.
N o t e
During actual CC operation, if a load change causes the voltage limit to be exceeded,
the power supply will automatically crossover to constant voltage mode at the preset
voltage limit and the output current will drop proportionately.
•
Remote interface operation:
VOLTage {<voltage>|MIN|MAX}
Set the voltage
CURRent {<current>|MIN|MAX}
Set the current
OUTPut ON
Enable the output
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the current.
Current
Voltage
1
Display
Limit
Display
Limit
On/Off
Output
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......