![Agilent Technologies E364XA Series User Manual Download Page 22](http://html1.mh-extra.com/html/agilent-technologies/e364xa-series/e364xa-series_user-manual_2868438022.webp)
Chapter 1 Quick Start
Line Voltage Conversion
21
1
Line Voltage Conversion
W a r n i n g
Shock Hazard
Operating personnel must not remove power supply covers.
Component replacement and internal adjustments must be made only by
qualified service personnel.
Line voltage conversion is accomplished by adjusting two components: the line
voltage selection switch and the power-line fuse on the rear panel.
1
Remove AC line power.
2
Remove the cover (Refer to General Disassembly on page 174).
3
Set two sections of the line voltage selector switch on the PC board for the
desired line voltage (See Figure 1-1 below).
4
See the next page to check the rating of the power-line fuse and replace with
the correct one if necessary.
5
Replace the cover and mark the power supply clearly with a tag or label
indicating the correct line voltage and fuse that is in use.
230V
100V
115V
Figure 1-1. Line Voltage Selector (set for 115Vac)
(TOP VIEW)
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......