![Agilent Technologies E364XA Series User Manual Download Page 155](http://html1.mh-extra.com/html/agilent-technologies/e364xa-series/e364xa-series_user-manual_2868438155.webp)
Chapter 7 Tutorial
Remote Programming
154
Remote Programming
During remote programming a constant-voltage regulated power supply is
called upon to change its output voltage rapidly. The most important factor
limiting the speed of output voltage change is the output capacitor and load
resistor.
Figure 7-7. Speed of Response - Programming Up (Full Load)
The equivalent circuit and the nature of the output voltage waveform when the
supply is being programmed upward are shown in Figure 7-8. When the new
output is programmed, the power supply regulator circuit senses that the
output is less than desired and turns on the series regulator to its maximum
value I
L
, the current limit or constant current setting.
This constant current I
L
charges the output capacitor C
O
and load resistor R
L
parallel. The output therefore rises exponentially with a time constant R
L
C
L
towards voltage level I
L
R
L
, a value higher than the new output voltage being
programmed.
When this exponential rise reaches the newly programmed voltage level, the
constant voltage amplifier resumes its normal regulating action and holds the
output constant. Thus, the rise time can be determined approximately using
the formula shown in Figure 7-8.
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......