Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
89
4
VOLTage:STEP {
<numeric value>
| DEFault}
Set the step size for voltage programming with the
VOLT UP
and
VOLT DOWN
commands. See the example below.
To set the step size to the minimum resolution, set the step size to ‘‘
DEFault
’’. The
minimum resolution of the step size is approximately 0.35 mV (E3646A), 1.14 mV
(E3647A), 0.38 mV (E3648A), and 1.14 mV (E3649A), respectively.
The immediate voltage level increases or decreases by the value of the step size. For
example, the output voltage will increase or decrease 10 mV if the step size is 0.01.
At
*RST
, the step size is the value of the minimum resolution.
VOLTage:STEP? [DEFault]
Return the value of the step size currently specified. The returned parameter is a
numeric value. ‘‘
DEFault
’’ gives the minimum resolution step size in unit of volts.
Example
The following program segments show how to use the
VOLT UP
or
VOLT DOWN
command to increase or decrease the output voltage with the
VOLT:STEP
command.
‘‘
VOLT:STEP 0.01
’’
Set the step size to 0.01 V
‘‘
VOLT UP
’’
Increase the output voltage
‘‘
VOLT:STEP 0.02
’’
Set the step size to 0.02 V
‘‘
VOLT DOWN
’’
Decrease the output voltage
VOLTage:TRIGgered {
<voltage>
| MINimum | MAXimum}
Program the pending triggered voltage level. The pending triggered voltage level is
a stored value that is transferred to the output terminals when a trigger occurs. A
pending triggered level is not affected by subsequent
VOLTage
commands.
VOLTage:TRIGgered? [MINimum | MAXimum]
Query the triggered voltage level presently programmed. If no triggered level is
programmed, the
VOLT
level is returned.
* For E3646A/48A models **For E3647A/49A models
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......