Chapter 3 Front-Panel Operation and Features
System-Related Operations
61
3
• Remote Interface Operation:
Use the following commands to store and recall power supply state.
*SAV {1|2|3|4|5}
*RCL {1|2|3|4|5}
To assign a name to a stored state to be recalled from the front panel, send the
following command. From the remote interface, you can only recall a stored state
using a number (1 through 5).
‘‘
MEM:STATE:NAME 1,
‘
P15V_TEST
’’’
Self-Test
A
power-on
self-test occurs automatically when you turn on the power supply. This
test assures you that the power supply is operational. This test does not perform the
extensive set of tests that are included as part of the complete self-test described below.
If the power-on self-test fails, the
ERROR
annunciator turns on.
A
complete
self-test performs a series of tests and takes approximately 2 seconds to
execute. If all tests pass, you can have a high confidence that the power supply is
operational.
If the
complete
self-test is successful, “
PASS
” is displayed on the front panel. If the
self-test fails, ‘‘
FAIL
’’ is displayed and the
ERROR
annunciator turns on. See the
Service Information
for instructions on returning the power supply to Agilent
Technologies for service.
• Front-panel operation:
To perform the complete front panel
self-test
, hold down the
key as you
turn on the power supply and
hold down
the key
until you hear a long beep
. The
self-test will begin when you release the key following the beep.
• Remote interface operation:
‘‘
*TST?
’’
Returns “0” if the complete self-test passes or “1” if it fails.
Error Conditions
When the front-panel
ERROR
annunciator turns on, one or more command syntax or
hardware errors have been detected. A record of up to 20 errors can be stored in the
power supply’s
error queue
. See ‘‘Error Messages’’ for more information starting on
Display
Limit
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......