Chapter 3 Front-Panel Operation and Features
Programming Overvoltage Protection
55
3
Programming Overvoltage Protection
Overvoltage protection guards the load against output voltages reaching values greater
than the programmed protection level. It is accomplished by shorting the output via
an internal SCR when the trip level is set to equal or greater than 3 volts, or by
programming the output to 1 volt when the trip level is set to less than 3 volts.
The following steps show how to set the OVP trip level, how to check OVP operation,
and how to clear overvoltage condition.
• Front-panel operation:
Setting the OVP Level and Enable the OVP Circuit
1 Turn on the power supply.
or
2 Select the output to be programmed.
3 Enter the OVP menu and set the desired trip level.
Use the knob and the resolution selection key
or
to set the desired trip
level.
Note
that you cannot set the trip levels to lower than 1.0 volt.
4 Enable the OVP circuit.
5 Exit the OVP menu.
If the OVP settings are not changed, “
NO CHANGE
” will be displayed. The power
supply will exit the OVP menu and the display will return to the meter mode. Check
that the
OVP1
and
OVP2
annunciators turn on if the OVP circuit is enabled.
LEVEL 2
2
.0V
(E3646A)
OVP ON
CHANGED
Power
1
2
Over
Voltage
<
>
Over
Voltage
Over
Voltage
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......