Chapter 3 Front-Panel Operation and Features
Constant Voltage Operation
42
Constant Voltage Operation
To set up the power supply for constant voltage (CV) operation, proceed as follows.
• Front-panel operation:
1 Connect a load to the output terminals.
With power-off, connect a load to the (+) and (-) output terminals.
2 Turn on the power supply.
The power supply will go into the
power-on / reset
state; the output is disabled (the
OFF
annunciator turns on); its low voltage range is selected (annunciator for the range
presently selected turns on, for example, the
8V
annunciator turns on for the E3646A
model); and the knob is selected for
voltage
control. At power-on, the output1 is
selected and the
OUT1
annunciator turns on.
Press
to operate the power supply in the high voltage range before proceeding
to the next step. The
20V
or
60V
annunciator turns on depending on which power
supply you are using.
3 Set the display to the limit mode.
Notice that the
Limit
annunciator flashes, indicating that the display is in the
limit
mode. When the display is in the
limit
mode, you can see the voltage and current limit
values of the power supply.
4 Adjust the knob for the desired
current
limit.
Check that the
Limit
annunciator still flashes. Set the knob for
current
control. The
flashing digit can be changed using the resolution selection keys and the flashing digit
can be adjusted by turning the knob. Adjust the knob to the desired current limit.
1
You can use the resolution selection keys to move the flashing digit to the right or
left when setting current.
In
constant voltage
mode, the voltage values between the meter and limit modes
are the same, but the current values are not. Moreover, if the display is in the
meter mode, you cannot see the change of current limit value when adjusting the
knob. We recommend that you should set the display to “limit” mode to see the
change of current limit value in the constant voltage mode whenever adjusting the
knob.
Power
High
Display
Limit
Current
Voltage
1
Current
Voltage
1
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......