Appendix Service Information
Troubleshooting Hints
175
Service
Information
Troubleshooting Hints
This section provides a brief check list of common failures. Before troubleshooting
or repairing the power supply, make sure that the failure is in the power supply rather
than any external connections. Also make sure that the power supply is accurately
calibrated. The power supply’s circuits allow troubleshooting and repair with basic
equipment such as a 6½-digital multimeter.
Unit Reports Errors 740 to 750
These errors may be produced if you accidentally turn off power of the unit during a
calibration or while changing a non-volatile state of the instrument. Recalibration or
resetting the state should clear the error. If the error persists, a hardware failure may
have occurred.
Unit Fails Self-Test
Verify that the correct power-line voltage setting is selected. Also, ensure that all
terminal connections are removed while the self-test is performed. Failure of the DAC
U20 on the PC board will cause many self-test failures.
Bias Supplies Problems
Check that the input to the voltage regulators of the bias supplies is at least
1 V greater than their output. Circuit failures can cause heavy loads of the bias supplies
which may pull down the regulator output voltages. Check the voltages of bias
supplies as tabulated below.
Table A-1 Bias Supplies Voltages
Some circuits produce their own local bias supplies from the main bias supplies. Be
sure to check that these local bias supplies are active. In particular, the ADC (analog-
to-digital converter), ac input, and front panel sections have local bias supplies.
Always check that the power supplies are free of ac oscillations using an oscilloscope.
Failure of bias supplies will cause many self-test failures.
Bias Supply
Minimum
Maximum
Check At
+5V Floating
+4.75 V
+5.25 V
U14 pin 2
-5.1V Floating
-4.75 V
-5.25 V
Anode of CR15
+15V Floating
+14.25 V
+15.75 V
Anode of CR139
-15V Floating
-14.25 V
-15.75 V
Cathode of CR140
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......