Chapter 8 Specifications
Supplemental Characteristics
166
Programming Language
SCPI (Standard Commands for Programmable Instruments)
State Storage Memory
Five (5) user-configurable stored states
Recommended Calibration Interval
1 year
Output Terminal Isolation (maximum, from chassis ground)
±60 Vdc when connecting shorting conductors without insulation between the (+)
output and the (+) sense terminals and between the (-) output and the (-) sense
terminals.
±240 Vdc when connecting insulated shorting conductors between the (+) output and
the (+) sense terminals and between the (-) output and the (-) sense terminals.
AC Input Ratings (selectable via two switches on PC board)
std
115 Vac ± 10%, 47 to 63 Hz
opt 0E3
230 Vac ± 10%, 47 to 63 Hz
opt 0E9
100 Vac ± 10%, 47 to 63 Hz
Maximum Input Power
Cooling
Fan cooled
Operating Temperature
0 to 40 °C for full rated output.
Storage Temperature
-20 to 70 °C for storage environment.
E3646A
E3647A
E3648A
E3649A
210 VA
210 VA
400 VA
330 VA
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......