Chapter 4 Remote Interface Reference
The SCPI Status Registers
108
The Standard Event register is cleared when:
• You execute the
*CLS
(clear status) command.
• You query the event register using the
*ESR?
(Event Status register) command.
For example, 28 (4 + 8 + 16) is returned when you have queried the status of the
Standard Event register, QYE, DDE, and EXE conditions have occurred.
The Standard Event Enable register is cleared when:
• You execute the
*ESE 0
command.
• You turn on the power and have previously configured the power supply using the
*PSC 1
command.
• The enable register will not be cleared at power-on if you have previously
configured the power supply using the
*PSC 0
command.
The Status Byte Register
The Status Byte summary register reports conditions from the other status registers.
Query data that is waiting in the power supply’s output buffer is immediately reported
through the “Message Available” bit (bit 4) of Status Byte register. Bits in the summary
register are not latched. Clearing an event register will clear the corresponding bits in
the Status Byte summary register. Reading all messages in the output buffer, including
any pending queries, will clear the message available bit.
Table 4-5. Bit Definitions – Status Byte Summary Register
Bit
Decimal
Value
Definition
0-2
Not Used
0
Always set to 0.
3
QUES
8
One or more bits are set in the questionable status
register (bits must be “enabled” in the enable register).
4
MAV
16
Data is available in the power supply output buffer.
5
ESB
32
One or more bits are set in the standard event register
(bits must be “enabled” in the enable register).
6
RQS
64
The power supply is requesting service (serial poll).
7
Not Used
0
Always set to 0.
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......