Absolute Measurement Fundamentals
5
Agilent E5505A User’s Guide
127
Tuning Range (PTR) necessary to provide a sufficient PLL bandwidth to
make the measurement.
Measurement options
If the observed level exceeded the small angle line at any point beyond the PLL
bandwidth set for the measurement, you need to consider one of the following
measurement options.
1
Evaluate your source using the noise data provided by the RF analyzer in
the procedure you just performed.
2
Increase the PTR if possible, to provide a sufficient PLL bandwidth to
suppress the noise. (For information on increasing the PTR, refer to
Changing the PTR in this section.)
3
Reduce the noise level of the signal sources.
4
Use the Discriminator technique to measure the phase noise level of your
source.
Figure 84
Requirements for noise exceeding small angle limit
e5505a_user_peak_tune_range.ai
rev2 10/24/03
Summary of Contents for E1420B
Page 24: ...24 Agilent E5505A User s Guide ...
Page 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Page 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Page 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Page 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Page 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Page 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Page 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Page 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Page 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Page 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Page 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...