Using the Oscilloscope
46
To set the oscilloscope to trigger on an edge
1
Press and release the
[Source]
key until the desired source LED is
illuminated.
You can choose any of the channels, the Aux, or Line as the source for an edge
trigger.
2
Press the
[Slope]
key until the desired slope LED is illuminated.
You can have an edge trigger on a rising or falling edge.
3
Press the
[Sweep]
key until the Trig’d LED is illuminated.
The oscilloscope will wait for the edge before initiating a sweep.
4
Turn the
[Level]
knob to adjust the voltage level at which the oscilloscope
will trigger.