386
Agilent 6000 Series Oscilloscopes Programmer's Reference
3
Commands by Subsystem
:TRIGger:GLITch:LEVel
(see
Command Syntax
:TRIGger:GLITch:LEVel <level_argument>
<level_argument> ::= <level>[, <source>]
<level> ::= .75 x full-scale voltage from center screen in NR3 format
for internal triggers
<level> ::= 2 V with probe attenuation at 1:1 in NR3 format for
external triggers
<level> ::= 6 V for digital channels (MSO models)
<source> ::= {CHANnel<n> | EXTernal} for DSO models
<source> ::= {CHANnel<n> | DIGital0,..,DIGital15} for MSO models
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger:GLITch:LEVel command sets the trigger level voltage for the
active pulse width trigger.
Query Syntax
:TRIGger:GLITch:LEVel?
The :TRIGger:GLITch:LEVel? query returns the trigger level of the current
pulse width trigger mode. If all channels are off, the query returns
"NONE."
Return Format
<level_argument><NL>