Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
323
:TIMebase:POSition
(see
Command Syntax
:TIMebase:POSition <pos>
<pos> ::= time in seconds from the trigger to the display reference
in NR3 format
The :TIMebase:POSition command sets the time interval between the
trigger event and the display reference point on the screen. The display
reference point is either left, right, or center and is set with the
:TIMebase:REFerence command. The maximum position value depends on
the time/division settings.
Query Syntax
:TIMebase:POSition?
The :TIMebase:POSition? query returns the current time from the trigger to
the display reference in seconds.
Return Format
<pos><NL>
<pos> ::= time in seconds from the trigger to the display reference
in NR3 format
See Also
•
"Introduction to :TIMebase Commands"
•
•
•
•
•
N O T E
This command is an alias for the :TIMebase:DELay command.