A Note About Backpressure
The ADM cartridge is calibrated with the standard flexible
tubing supplied with the meter at the time of purchase.
Replacement tubes should be of equal internal diameter and
length. Be aware that using tubing that increases or decreases
the total volume of gas in the tubing will affect accuracy. Tubing
of greater volume will result in slightly lower flow readings. This
is because the transducer creates an ever-so-small
backpressure that compresses the gas slightly.
Technical Support
www.agilent.com/chem
).
Other Features
•
Automatic power off
•
Split ratio mode with continuous split flow rate reading
•
NIST traceable calibration
•
Compatible with dry gases
•
Volumetric and mass flow measurements
•
USB communication
•
Replaceable flow meter cartridge - individually computer
calibrated