Chapter 6: Maintenance and troubleshooting
BioCel User Guide
86
Routine maintenance
About routine
maintenance
The BioCel does not require a lot of routine maintenance.
In general, practice good housekeeping by cleaning up spills and
following the clean-up procedures.
Once a month, check that the:
❑
Robot gripper pads are not torn, cracked or otherwise worn
❑
Overhead doors remain at the correct height
❑
Liquid handling tubing is not torn, cracked or discolored
❑
Air filter is clean
❑
Main air pressure reads 100–110 psi (0.62–0.69 MPa)
❑
Table top is free of debris, such as pieces of chipped plates and
microtubes
Checking the air
filter for
replacement
If the house air supplied to the BioCel is clean, the air filter should rarely
need to be changed.
To check the air filter for replacement:
1.
Open the side door behind the air panel.
2.
Look at the front of the air filter container.
If there are any oil droplets or general dirt visible on the filter, the
filter should be changed. Contact the Velocity11 Technical Support
for instructions.
Device maintenance
For information about routine maintenance of the individual devices,
see the relevant device documentation.
Replacing
fluorescent tube
lights
The interior of the BioCel has standard type fluorescent tube lights that
occasionally need to be replaced. We recommend that two people work
together to replace a light.
To replace a fluorescent tube light:
1.
If your BioCel has a light switch, turn off the lights.
2.
At the light itself, unplug the power cord that goes to the light.
3.
Remove the screws securing the light cover.
4.
With one person working at each end of the light, remove the light
cover by sliding the cover up and pulling out at the bottom.
5.
Remove the lamp.
6.
Install a new, identical lamp.
7.
Replace the cover.
Summary of Contents for BioCel 1200
Page 1: ...Agilent Technologies BioCel System User Guide...
Page 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division...
Page 18: ...Chapter 1 Introduction BioCel User Guide 12...
Page 64: ...Chapter 3 Preparing for a run BioCel User Guide 58...
Page 74: ...Chapter 4 Performing a run BioCel User Guide 68...
Page 100: ...Chapter 6 Maintenance and troubleshooting BioCel User Guide 94...
Page 104: ...Index BioCel User Guide 98...
Page 105: ......