![Agilent Technologies BenchCel User Manual Download Page 96](http://html1.mh-extra.com/html/agilent-technologies/benchcel/benchcel_user-manual_2867698096.webp)
4 Setting up the BenchCel device
Setting and managing teachpoints
86
BenchCel Microplate Handler User Guide
Determining the teachpoint coordinates
WARNING
To prevent potential injury, make sure no one else can issue commands at the
controlling computer while you are manually determining the teachpoint coordinates.
To determine the coordinates of the pickup and placement location:
1
Place two or three spare microplates in a labware rack, and then install the rack on
the BenchCel Microplate Handler. For instructions, see
“Installing and uninstalling the labware racks” on
.
2
Open BenchCel Diagnostics for the BenchCel device that you are setting up.
3
On the
Profiles
page, look in the
Teachpoint File
area to make sure the correct
teachpoint file is loaded. You will be adding new teachpoints to this file.
4
In the
Controls
tab, select the labware definition from the
Labware
list.