2-27
Using the Multi-Wavelength Meter
Calibrating Measurements
To enter the elevation
1
Press the
Setup
key.
2
Press the
MORE
softkey.
3
Press the
CAL
softkey.
4
Press
ELEV
.
5
Use the
and
softkeys to enter the elevation in meters. Entries jump in
500 meter steps from 0 m to 5000 m.
In order for the Agilent 86120B to meet its published specifications, the eleva-
tion value selected with the softkeys must be within 250 meters of the actual
elevation.
6
Press
RETURN
to complete the entry.
To select the medium for light
1
Press the
Setup
key.
2
Press the
MORE
softkey.
3
Press the
CAL
softkey, and make the following selection:
• Press
VACUUM
for wavelengths in a vacuum.
• Press
STD AIR
for wavelengths in standard air.
4
Press
RETURN
to complete the entry.
Converting feet to meters
If you know your elevation in feet, you can convert this value to meters by using the fol-
lowing equation:
m
ft
3.281
---------------
=
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......